ssa3021Last year  in the laboratory where I work we bought two Siglent SSA 3021X 2.1 GHz Spectrum Analyzer for generic/every day use. It’s a nice cheap object with all the functions of a modern digital instrument: colors, memory and math for traces, ethernet access, usb drive support etc.
After a while I found some bug in the software (v 7.03), as example if a screenshot image is saved in the internal memory with spaces in the filename, then it’s impossible to delete….
So I googled for a firmware update and casually found a thread on EEVBlog’s Forum where a guy shared the root’s user/password to log in the instrument with a telnet client. :-)

The big brother of the SSA3021X is the SSA3032X 3.2 GHz SA that has the same hardware of the SSA3021X. The instrument works on Linux, Busybox is running, and it’s an easy job to edit a configuration file and upgrade the SSA3021X to SS3032X with all the options enabled (tracking generator included) and extra BW filters 1Hz and 3MHz (!!!)

This works with installed firmware version 7.07, probably Siglent has corrected this bug on the newer firmware releases, but for sure I don’t need further updates on these units…

ssa3021x_to_3032x